At Teleflex, we provide a growing portfolio of innovation that can help you reduce costs, drive positive clinical outcomes, and increase patient satisfaction. We are committed to supporting our partnership with your facility through our Weck™ 360 Service Program where we offer educational tools on Weck™ applier care and maintenance.
All appliers should be cleaned and sterilized per instructions provided in the IFUs
Follow hospital or national guidelines on cleaning of appliers.
Opening the box lock:
Weck™ Hem-o-lok™ Take-Apart Endoscopic Clip Appliers help facilitate access to organic matter and blood that can accumulate in areas that may be difficult to reach during cleaning. They disassemble into three pieces for thorough cleaning to help your facility maintain standards for care and maintenance of your surgical instrumentation.
Lubrication is essential every time instruments are processed. Do not use mineral oil, petroleum, or silicone-based products. Lubricate joints using a non-silicone, water-based surgical-grade lubricant.
Please examine endoscopic and open appliers and removers for potential damage before use in surgery. Pay particular attention to the jaws. Damaged or misaligned jaws may not allow clips to close acceptably for occlusion of intended structure. Damaged remover jaws may not allow for clip removal. If repair is required, clean, sterilize and return damaged instrument to Teleflex Medical.
Alignment:
Check the following components of your open clip appliers:
Spring:
Flag Condition:
Your local Teleflex Surgical sales representative is your primary resource for support. Additional sales resources, including our sales trainers and regional management, are available for additional consultation as needed.
Weck Ligating Clips are contraindicated for use as a fallopian contraceptive tubal occlusion device and
contraindicated for use in ligating the renal artery during laparoscopic donor nephrectomies.
Teleflex, the Teleflex logo, Hem-o-lok, Horizon, and Weck are trademarks or registered trademarks of
Teleflex Incorporated or its affiliates, in the U.S. and/or other countries. Other names may be
trademarks of their respective owners.
Information in this material is not a substitute for the product Instructions for Use. Not all products
may be available in all countries. Please contact your local representative.
© Teleflex
Incorporated. All rights reserved. Revised 02/2022. MC-006240 Rev 0.1